-
Notifications
You must be signed in to change notification settings - Fork 6
Identifying Lipid Chromatographic Peaks
Peak Finder requires .csv peak tables generated from either Compound Discoverer or mzMine 2 and .csv MS/MS result files generated by LipiDex and identifies lipid chromatographic peaks. The resulting peak table is extensively filtered to generate high-confidence lipid identifications with minimal requires hand-annotation. Results are exported as .csv files to the same directory as the peak table.
-
Open Peak Finder
-
Select the Peak Table Type
-
LipiDex supports peak tables from Compound Discoverer or MZmine 2 in .csv format. Instructions for generating these peak tables can be found using the following links:
-
-
Upload Peak Tables
-
For each field, select “Add” and locate the applicable peak tables.
-
For Compound Discover, the required peak tables are an aligned and an unaligned table. For MzMine 2, a peak table for both positive and negative polarity are required.
-
Loading Peak Tables for Compound Discoverer
Loading Peak Tables for MZmine 2
-
Select MS/MS Result Files
-
Select the “Load Files” button
-
In the “Results Uploader” window, select the appropriate data acquisition type.
-
Select “Add Files” to locate the Spectrum Searcher .csv result files
-
If “Separate Polarity Analysis” is selected, edit the “File ID” field in the table. The positive and negative polarity experiments to be merged should share the same File ID number.
-
Select “Upload”
-
-
Update MS/MS Parameters as Needed
-
The “Min. Lipid Spectral Purity” parameter specifies the minimum spectral purity needed to annotate a lipid at the molecular composition level (PC 16:1_18:1) rather than at the sum composition level (PC 34:2).
-
The “Min. MS2 Search Dot Product” parameter specifies the minimum spectral similarity score needed to use an MS/MS identification.
-
The “Min. MS2 Search Rev. Dot Product” parameter specifies the minimum reverse spectral similarity score needed to use an MS/MS identification.
-
The “FWHM Window Multiplier” parameter specifies the maximum allowed retention difference between the apex of the chromatographic peak and the MS/MS spectra in terms of the FWHM of the chromatographic peak.
-
The “Max. Mass Difference” parameter specifies the maximum relative mass difference (ppm) allowed to associate a lipid identification with a chromatographic peak.
-
-
Update Result Filtering Parameters as Needed
-
Select possible adducts for feature merging
-
Select “Configure”
-
Add all adducts observed in the LC-MS/MS experiment using the same formatting described in the “Creating and Editing a Lipid Library” section of this document.
-
The “Max. RT M.A.D. Factor” parameter specifies the maximum allowed retention time difference between a lipid identification and the all other identified lipids of the same class in terms of multiples of the median absolute retention time deviation of the lipid class.
-
The “Feature Found in n Files” parameter specifies the minimum number of times the specific feature was identified to be included in the final peak table.
-
-
Identify Lipid Chromatographic Peaks
-
Select the “Identify Chromatographic Peaks” button
-
Upon completion all result files are written to the same directory as the peak tables
-
The filtered peak table is written to Final_Results.csv
-
The peak table which includes all peaks which have been filtered is written to Unfiltered_Results.csv
-
Sample-specific metrics are written to Sample_Information.csv
-
Tutorials